Syringe pump KSP–A100 is a light weight, portable device designed for low-volume infusion. Equipped with LCD screen to display parameters like injection state, volume, mode, speed, syringe size, alarms, dosage, occlusion status, etc. for enhanced clinical efficiency. Inbuilt software based control system delivers higher levels of precision, safety and allows remote control mechanism to improve standardization, accuracy analysis of infusion data and reduce free–flow risks.

Specification

KVO Rate 0.1 – 5ml/h (steps : 0.1ml/h)
Injection Compliance Adaptable for 5 syringe sizes
Preset Volume Range 0.1 ml – 1999.9 ml
Occlusion System 3-level alarming system
Mechanical Precision ≤ ±2 %
Injection Rate Range
50 ml Syringe0.1 ml – 999.9 ml/h (step: 0.1 ml/h)
1000 ml – 1800 ml/h (step: 1 ml/h)
30 ml Syringe0.1 ml – 900 ml/h (step: 0.1 ml/h)
20 ml Syringe0.1 ml – 600 ml/h (step: 0.1 ml/h)
10 ml Syringe0.1 ml – 300 ml/h (step: 0.1 ml/h)
5 ml Syringe0.1 ml – 150 ml/h (step: 0.1 ml/h)
Purge Rate
50 ml Syringe1800 ml/h (Bolus rate: 1200 ml/h)
30 ml Syringe900 ml/h (Bolus rate: 600 ml/h)
20 ml Syringe600 ml/h (Bolus rate: 400 ml/h)
10 ml Syringe300 ml/h (Bolus rate: 200 ml/h)
5 ml Syringe150 ml/h (Bolus rate: 100 ml/h)
Occlusion Alarm Threshold
High800 mmHg ± 200 mmHg (106.7 kPa ± 26.7 kPa)
Medium500 mmHg ± 100 mmHg (66.7 kPa ± 13.3 kPa)
Low300 mmHg ± 100 mmHg (40.7 kPa ± 13.3 kPa)
Fuse F1AL/250V (2 nos.)
Power Supply 100 V – 220 V, 50 Hz
Battery Built–in Rechargeable Li–ion battery (11.1V)
Capacity ≥ 2000mAh ; 10 hours of charging
Works at 5ml/h after 4 hours of charging
Alarm Near Injection finished, Preset dose injection finished, Occlusion, Syringe off, Syringe dislocated, Low Battery, On AC Supply.
Safety Class I Type BF
Operation
Ambient Temperature5°C to 40°C
Relative Humidity20% – 90%
Atmospheric Pressure86 kPa – 106 kPa
Storage
Ambient Temperature–30°C to 55°C
Relative Humidity≤ 95%
Atmospheric Pressure50 kPa – 106 kPa
Dimensions 282 x 210 x 148 mm